Technical specs:
Capacity 0.8 g / 3.1 g
Readability 1 µg / 10 µg
Pan Size 16 mm / 0.63″ Dia.
Repeatability (at nominal load) 0.001 mg / 0.006 mg
Repeatability (at low load) 0.0008 mg (0.2 g) / 0.005 mg (0.2 g)
Linearity 0.004 mg
Eccentric load deviation (test load)
According to OIML R76 0.005 mg (2 g)
Repeatability (sd) 0.0005 mg + 1.2 x (10-7) · R_gr
Differential linearity deviation (sd) ?êÜ2 x (10-12)g · R_nt
Differential eccentric load deviation (sd) 1.2 x (10-6) · R_nt
Sensitivity offset (sd)
(In the temperature range 10 – 30°C) 3 x (10-6) · R_nt
Minimum weight (@ U=1 %, 2 sd) 0.1 mg + 2.4 x (10-6) · R_gr
Settling time (standard range) < 6 Sec
Settling time (fine range) < 10 Sec
METTLER MX5 ANALYTICAL MICROBALANCE
Automated weighing for maximum savings
Increase your productivity in the weighing of short and narrow tubular samples with METTLER TOLEDO MX5 Automated-S high-speed microbalance solutions.
To receive support or a quotation, please click on the appropriate button. Your message is routed directly to your local sales representative who will provide you with the requested information.
Tips & Tricks
Weighing the Right Way
AntiStatic Kit – Discharge Your Samples
Datasheets
MX5 Automated-S/UMX2 Automated-S
Declaration of conformity
Declaration of conformity
Certificates
OIML certificate of conformity (R76/1992-NL1-05.06)
EC type-approval certificate (T6361)
Manuals
Operating Instructions AX and MX/UMX Balances
Supplement to the operating instructions MX5-S and UMX2-S automated for MX and UMX balances
Operating instructions Excellence MT-SICS
Free Standard Operating Procedures (SOP’s) for Balance Straightforward Testing
There are no reviews yet.